STEL 3 – A Product of Linear Chemicals
- 3part Diff
- 20 Parameters
- 3 Histogram for WBC, RBC and PLT
- Throughput: 60tests/hr.
- Counting modes: Whole Blood, Prediluted and Peripheral
- Sample Vol.: 9.8 μl Total Blood 20 μl Pre-Diluted
TEST | SPECIFICATIONS | RESULTS |
---|---|---|
Blank counting | Result of a measurement using Hematology Analyzer, test value should be: White blood cell count(WBC)≤0.3*109/L, Red blood cell count(RBC) ≤0.05*1012/L, Platelet count (PLT) ≤10*109/L, Hemoglobinometry (HGB) ≤2g/L. |
WBC:0 0 0 RBC:0 0 0 PLT:0 0 0 HGB:0 0 0 |
Wavelength of absorption peak |
Wavelength of absorption peak of haemoglobin ramification after erythrolysis by Lyse should not exceed ±10nm of nominal value | 537.6 |
Absorbance Value |
At wavelength of 750nm, absorbance value of fresh blood after erythrolysis should not exceed 0.012. |
0.010 |
WBC Histogram | Analyze normal person’s fresh blood using hematology analyzer. The WBC histogram should meet the following requirements: 1. Include two group peaks of small and large cells. 2. Should be in accordance with the shape of peak and range of peak position of WBC partition histogram of related hematology analyzer. 3. There should not be interference peak within 35fl |
pass |
Reviews